Combined-languages-d1 - Flipbook - Page 38
USER'S MANUAL
EVOLUTION VETERINARY DENTAL UNITS
38 � EN
Cleaning and disinfecting surfaces
» Regularly clean and disinfect surface and thread of the
bottle and bottle receptacle, e.g. with FD 366 sensitive
top wipes.
6.3 Once or twice a week/break-time
preservation
The following are required for cleaning:
» Non-foaming disinfectant/cleaning agent that is
Fig.38 -Bottles Expiry date on bottle
iM3 recommends:
9 For continual disinfection of procedural water and
water supply lines: Alpron* solution
9 For cleaning and disinfection of waterways:
Bilpron* solution
9 For cleaning and disinfection of bottles:
BC-San 100* solution
9
For cleaning and disinfection of external surfaces,
bottles and threads: FD 366 sensitive top wipes
Disinfection of sensitive surfaces and FD 366
sensitive Disinfection of sensitive surfaces, ready-touse solution**
Only these products have been tested by iM3.
* Alpron, Bilpron and BC-San 100 are products from Alpron Medical GmbH.
** FD 366 sensitive top wipes and FD 366 sensitive Disinfection of sensitive
surfaces, ready-to-use solution, are products from DÜRR DENTAL SE.
6.1 Procedural water disinfection
Drinking water can be mixed with Alpron for use on
patients. This ensures consistent water quality.
» Detailed information can be found in the Alpron
operating instructions.
6.2 Daily after the end of treatment
The bottle contents must be changed every day,
otherwise there is a risk of contamination.
The following are required for cleaning:
» Non-foaming disinfectant/cleaning agent that is
compatible with the materials.
» Clean the bottle with hot water (≤ 90 °C) and a cleaning
brush.
compatible with the materials.
» Rinse the bottle with hot water.
» Clean the bottle with a cleaning brush or in the washerdisinfector.
» Fill the bottle with Bilpron according to the
manufacturer’s instructions and flush all the lines.
Detailed information can be found in the Bilpron
operating instructions.
» Fill the empty bottle with BC-San 100 according to the
manufacturer’s instructions and close with a screw cap, if
available.
A bottle that is filled with BC-San 100 solution
must not be connected.
Detailed information can be found in the BC-San 100