Catalogue - Endoscopic - Flipbook - Page 60
SHAVER Arthroscopic shaver
Article number
SH6001000
Strengths
(Shaver Set: unit, hand piece, footswitch
·
·
·
·
·
·
·
Easy, efficient and reliable cleaning and desinfection process resulting from unique head design
Sophisticated oscillation modes to meet requirements of the most demanding operator
Smooth and stable work in all range of rotational speed
Singleuse and reusable shaver blades with uniquely designed hub eliminating problematic articulation effect
Large and clearly visible user interface facilitates reading of essential parameters of the device
Remote interface enables presentation of significant parameters on the medical monitor
Convenient manipulation by means of the manually controlled buttons available on the shaver handpiece
Shaver control unit
Dimensions:
· 250mm x 330mm x 96mm (DxWxH), 4,8 kg
System:
· Power supply: 100240 V AC; 50/60 Hz
Power consumption:
· 65 VA
Class of protection:
· Class I
Ambient temperature / Max. relative humidity: · During operation: +10˚C to +40˚C / 70%
· During storage and transport: 20˚C to +45˚C / 70%
CE Standard
· EN606011; EN6060112
SH6001001
Shaver hand piece
Dimensions:
System:
Shaver blade chuck:
Speed range:
Oscillation modes:
Cable lenght:
Sterilization method:
Ambient temperature / Max. relative humidity:
Class of protection:
· Length: 164 mm; 325 g
· Power supply: 100240 V AC; 50/60 Hz
· Automatic / Barracuda type
· FWD / REV: 500 8000 RPM (+/10%)
· OSC: 500 3500 RPM (+/10%)
· 9 different modes
· 3,2 m
· Steam sterilization 134 ˚C
· During operation: +10˚C to +40˚C / 100%
· During storage and transport: 20˚C to +45˚C / 90%
· Application part: BF type
SH6001002
Shaver footswitch
Dimensions:
Cable lenght:
Casing savety standard:
Functions:
215mm x 280mm x 28mm (DxWxH), 2,1 kg
3,2 m
IP X7
Speed up, speed down, start/stop, window size, direction
Ambient temperature / Max. relative humidity:
· During operation: +10˚C to +40˚C / 100%
· During storage and transport: 20˚C to +45˚C / 90%
55