Inventis Catalogue - Flipbook - Page 33
Trumpet
REM system and PC controlled
audiometer
Real Ear Measurement system
Real Measurement standard
Stimuli (REM)
IEC 61669, ANSI S3.46.
Applicable standards
White noise, Pink noise, ICRA,
ISTS, Real Speech, mixed
environment sounds
125 Hz – 12 kHz
Intensity range
50 – 90 dB SPL
Measure range
probe mic 40 - 130 dB SPL
ref mic 40 - 110 dB SPL
Available tests
1/3, 1/24 octave
Available tests (stand. mode)
Available tests (custom mode)
Sweep & Broadband
frequency measure,
THD, Battery drain,
Input/Output,
Attack/Release
REUR / REUG, REAR / REAG,
REIG, REOR / REOG, Advanced,
Live speech, RECD (opt.)
RECD
2A
Output
AC, BC, Speech
Stimuli
pure tone, warble tone
Speech inputs
Special tests
EN 60118-0, EN 6011-7, ANSI S3.22
OSPL90, Full On Gain, I/O,
Attack/Release, RTG,
Frequency response,
Equivalent Input Noise,
THD, Battery Drain, Telecoil
Audiometer
Type EN 60645-1, ANSI S3.6
Hearing Instrument Test (HIT) box
Hearing Aid Analyzer
Frequency range
Frequency Resolution
Type of device
Frequency range
Frequency resolution
Yes (if combined
with Trumpet REM)
125 Hz – 10 kHz
1/3, 1/6, 1/12, 1/24 octave
USB, Mic
Stimulation intensity range
40 - 100 dB SPL
Master Hearing Aid,
QuickSIN™ (opt.)
Coupler measurement range
40 - 145 dB
Max. output intensities
Attenuation
@ 1kHz | › 30dB
AC
-10 to 120 dB HL
Coupler type
2cc (ITE, RIC and BTE adapters)
BC
-10 to 80 dB HL
Telecoil drive
10 - 100 mA/m
FF output (amplified)
Connection to PC
Controlling software
Noah compatibility
Dimensions (WxDxH)
Weight
Yes (external or built-in speaker)
USB (driverless)
Maestro (included)
Yes
15.5 x 10 x 24.5 cm / 6.1 x 3.9 x 9,7 in
1.5 kg / 3.5 lbs
Battery simulator
Connection to PC
Controlling software
Noah compatibility
Dimensions (WxDxH)
Weight
Standard 10A, 13, 312, 675
USB (driverless)
Maestro (included)
Yes
32 x 33 x 22 cm / 12.6 x 13 x 8.7 in
7.5 kg / 16.5 lbs
33
TECHNICAL DATA
Type of device
Drum